It is important to prevent water getting into the sample chamber because
this leads to incorrect measurement results.
8.
Dirt in the transparent sample chamber leads to incorrect measurement
results. The light entry surfaces should be checked at regular intervals and
cleaned as necessary.
Standard spectacle cleaning cloths and cotton buds are suitable for
cleaning.
9.
Pronounced differences in temperature between the photometer and the
surroundings can lead to incorrect measurements, e.g. due to the forma‐
tion of condensation in the sample chamber and on the cuvette.
10.
Protect the device from direct sunlight.
Fig. 2: Correct filling of the cuvette: Left = correct / right = incorrect
11.
Fill the cuvette as shown in Fig. 2.
12.
Insert the reagent tablets directly from the blister pack into the water
sample, without touching them with your fingers.
13.
After use immediately close the dropping bottles containing the liquid
reagents by closing them with the screw caps of the same colour.
14.
You must observe the sequence for the addition of reagents without fail.