Calibrating Measuring Rod (Hcal); Restoring Factory Settings (Reset) - Seca 787 Instructions For Use And Warranty

Table of Contents

Advertisement

Calibrating measuring rod (HCAL)

150,0 cm

Restoring factory settings (rESEt)

36 •
4. Confirm your selection.
You exit the menu automatically.
If the value displayed in the display deviates from the value shown on read-off
edge 2, calibrate the measuring rod.
1. Select the
2. Confirm your selection.
SEt
3. Set the measuring rod to 150 cm.
4. Confirm the setting.
The measuring rod is calibrated.
You exit the menu automatically.
.
If you restore the factory settings (variant-dependent), the settings for the
following functions will be reset:
Units of height and weight are reset to metric units.
1. Select the
2. Confirm your selection.
You exit the menu automatically.
3. Switch off the measuring station.
Factory settings are restored and are available when the measuring
station is switched back on.
item from the menu.
HCAL
and
are displayed.
150.0 cm
– Autohold (
)
AHOLd
– Permanent display of weight value (
– Filtering (
)
FIL
– Standby time (
)
AOFF
– Switching units of measurement (
NOTE
only: If you restore the factory settings, the WiFi settings will
seca 797
be retained.
item from the menu.
rESEt
)
HOLd
)