Problem
The Workflow LED illuminates
in red, an error code
(Er:8:[No.]) is displayed
Er:8:94 is displayed
Er:8:95 is displayed
The rF menu item is not visible
(seca 797)
In the rF menu, only the SYS
menu item is visible
Id:Er is displayed

9. SERVICING

10.TECHNICAL DATA

10.1 General technical data

Dimensions
• Depth
• Width
• Height (measuring rod not extended)
Weight of measuring station
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
40 •
Cause
Errors occurred during the measurement
procedure or data transmission
An error occurred during the measurement
procedure: IDs were scanned at the wrong
time
Hardware fault
The internal interface module of the device
is defective
The WiFi function is deactivated
The ID of the user or the patient has not
been found in the EMR system
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
General technical data
Remedy
• Contact your administrator or hospital
technician
• Cancel measurement procedure
• Repeat measurement procedure and
observe proper sequence
• If the problem persists, contact your
administrator or hospital technician
• Contact your administrator or hospital
technician
• Inform seca Service
• Inform seca Service
• Activate WiFi function
• Rescan ID
• If the problem persists, contact your
administrator or hospital technician
Approx. 13.1 kg
+10 °C to +40 °C / +50 °F to 104 °F
700 – 1060 hPa
30 % – 80 %, no condensation
-10 °C to +65 °C / +14 °F to 149 °F
700 – 1060 hPa
0 % – 95 %, no condensation