4.3.5

Path resistance tests

Perform the following steps to verify that the path resistance
of each channel is with specified values.
1. Connect the HI and LO terminals of all nine channel in-
puts together using #22AWG solid copper jumper wires
(see Figure 4-2).
4-4
Reference
Thermometer
Input
SENSE
INPUT
Ω 4 WIRE
HI
350V
1100V
PEAK
PEAK
2001 MULTIMETER
LO
500V
INPUTS
PEAK
Ω2
Ω4
DCI
ACI
FREQ
TEMP
F
R
RANGE
AUTO
FRONT/REAR
STORE RECALL
FILTER
MATH
2A 250V
AMPS
RANGE
CAL
CHAN
SCAN
CONFIG
MENU
EXIT
ENTER
Copper Wires
Model 2001 #2
(Measure DCV)
Model 2001-TCSCAN installed in
Model 2001#1 (Not shown)
HI
Input
LO
OUT A HI
OUT A LO
Make sure that all jumper wires are clean
and free of oxides.
2. Connect the output cables to OUT A HI and LO, as
shown in Figure 4-2.
3. With the power off, install the scanner card in the Model
2001 Multimeter.