Keithley 2001-TCSCAN Instruction Manual page 56

Scanner card
Hide thumbs Also See for 2001-TCSCAN:
Table of Contents

Advertisement

Figure 3-5
Combining 2-pole and 4-pole switching
Combining 2-pole and 4-pole switching
The Model 2001-TCSCAN can combine 2- and 4-pole
switching in various combinations. For example, you might
require 4-pole switching for only two resistors, while the re-
maining resistors could be tested using conventional 2-pole
switching.
Out A
Input HI
HI
Input LO
LO
DMM
Out B
Sense HI
HI
Sense LO
LO
2001-TCSCAN
CH 2
CH 3
CH 4
CH 5
CH 6
CH 7
CH 8
CH 9
CH 10
Figure 3-5 shows a test configuration using mixed 2-pole and
4-pole switching. Four-pole switching is used only for the
device connected to channels 2 and 7. Five more resistors are
tested using 2-pole switching. In order to use this
configuration, configure the channels for Ω2 and Ω4
functions as appropriate.
Operation
DUT's (6)
3-27