Important Safety
Information
4

Servicing Information

Refer all servicing to qualified service personnel. Servicing is required when
the apparatus has been damage in any way, such as when an antenna cable or
Ethernet cable or connector is damaged, the unit has dropped from height, the
sensor's casing has been opened, or the device does not operate normally.

Regulatory Information

All Keysight Technologies Inc. devices are designed to be compliant with rules
and regulations in locations they are sold and will be labeled as required. Any
changes or modifications to Keysight equipment, not expressly approved by
Keysight, could void the user's authority to operate the equipment.
When Keysight devices are professionally installed, the Radio Frequency
Output Power will not exceed the maximum allowable limit for the country of
operation.
Unauthorized modifications or attachments could cause equipment damage
and may violate regulations.
Radio Compliance
Complies with the essential requirements of the European
(GPS)
Radio Equipment Directive as well as current editions of the
following standards (dates and editions are cited in the
Declaration of Conformity):
EMC
Complies with current editions of the following standards
compliance
(dates and editions are cited in the Declaration of
Conformity):
Note 1: Compliance with Radiated Emissions limits requires the use
of a double-shielded, Cat 7 Ethernet cable with the weatherproof
boot. (See page 19 in the Installation Guide for instructions.)
Note 2: This is a sensitive measurement apparatus by design and
may have some additional residual responses when exposed to
ambient continuous electromagnetic phenomenon in the range of
150 kHz-2.7 GH when tested per IEC 61000-4-3 or IEC 61000-4-6.
Based on a small sample size, performance loss was observed, up to
-40dBm compared to a measured baseline of -120dBm.
Safety
IEC 61010-1:2010 / EN 61010-1:2010 (3
compliance
ITU-R
ITU-R SM.377 between 20-6000MHz (frequency and
Compliance
frequency offset measurements)
Acoustic statement:
Acoustic noise emission
(European Machinery
LpA <70 dB
Directive)
Operator position
Normal operation mode per ISO 7779
To find a current Declaration of Conformity for a specific Keysight product, go
to:
http://regulations.about.keysight.com/DoC/search.htm
-
EN 303 413
-
EN 300 440
· IEC/EN 61326, CISPR Pub 11 Group 1, Class B
· AS/NZS CISPR 11
· ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du
Canada.
-
ITU-R SM.378 (field strength measurements)
-
ITU-R SM.328 (bandwidth measurements)