BEAM ALIGNMENT
PROCEDURE
Figure 2-14. Service Module
Switch Location
2-18
NOTE
Alignment can not be done unless the power is applied to the stack mounted
service module. The control unit does not have to be connected or powered to
perform an alignment. For optimal alignment accuracy the stack should be at
normal operating temperature. Alignment should be completed before
instrument calibration.
Service module switches (Figure 2-14) should be in the normal operating
positions given below.
• Operate/Zero - Operate
• Operate/Span - Operate
• Normal/Test - Normal
2
• Opacity/T
- Opacity
OPERATE
ZERO
SPAN
NORMAL
EXTERNAL
METER (mA)
TEST
OPACITY
00.3
T
2
0 100%
Model:
Serial:
Instruction Manual
IM-105-4000, Rev 1.3
October 2008