6 Patient Alarms and INOPs
Alarm Message
*** EXTREME BRADY
***BRADY xxx < yyy
*** EXTREME TACHY
***TACHY xxx > yy
** FHR1 HIGH
** FHR2 HIGH
** FHR3 HIGH
**FHR1 xxx>yyy
**FHR2 xxx>yyy
**FHR3 xxx>yyy
** FHR1 LOW
** FHR2 LOW
** FHR3 LOW
**FHR1 xxx<yyy
**FHR2 xxx<yyy
**FHR3 xxx<yyy
** HR HIGH
**HR xxx>yyy
** HR LOW
**HR xxx<yyy
** NBPs HIGH
** NBPd HIGH
** NBPm HIGH
** NBPs LOW
** NBPd LOW
** NBPm LOW
74
From
Condition
MECG
The maternal heart rate obtained
from the maternal ECG has fallen
below the extreme bradycardia limit.
xxx denotes the lowest measured
value, and yyy is the extreme
bradycardia limit.
MECG
The maternal heart rate obtained
from the maternal ECG has risen
above the extreme tachycardia limit.
xxx denotes the highest measured
value, and yy is the extreme
tachycardia limit.
FHR (ultrasound) The fetal heart rate obtained from
ultrasound has risen above the high
alarm limit.
xxx denotes the highest measured
value, and yyy is the high alarm limit.
FHR (ultrasound) The fetal heart rate obtained from
ultrasound has fallen below the low
alarm limit.
xxx denotes the lowest measured
value, and yyy is the low alarm limit.
MECG
The maternal heart rate obtained
from the maternal ECG has risen
above the high alarm limit.
xxx denotes the highest measured
value, and yyy is the high alarm limit.
MECG
The maternal heart rate obtained
from the maternal ECG has fallen
below the low alarm limit.
xxx denotes the lowest measured
value, and yyy is the low alarm limit.
Noninvasive
The measured noninvasive blood
blood pressure
pressure value is above the high
alarm limit.
s
d
,
, or
whether the systolic, diastolic or
mean pressure has crossed the limit.
Noninvasive
The measured noninvasive blood
blood pressure
pressure value is below the low alarm
limit.
s
d
,
, or
whether the systolic, diastolic or
mean pressure has crossed the limit.
m
after the label indicates
m
after the label indicates
Indication
Numeric flashes, red alarm
message, alarm tone.
Numeric flashes, red alarm
message, alarm tone.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.
Numeric flashes and high
limit is highlighted, yellow
alarm message, alarm tone.
Numeric flashes and low limit
is highlighted, yellow alarm
message, alarm tone.