Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user must ensure
that it is used in the appropriate environment as described below.
Table 2 - Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test Level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV differential
IEC 61000-4-5
mode
± 2 kV common
mode
Voltage dips, short
<5% U
interruptions and
in U
T
voltage variations
40% U
on power supply
U
) for 5 cycles
T
input lines
70% U
IEC 61000-4-11
(30% dip in U
25 cycles
< 5% U
(> 95% dip in U
5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Key: U
is the AC mains voltage prior to application of the test level.
T
Finding Recommended Separation Distances
In the following table, P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer, and d is the recommended separation distance in meters (m).
Portable and mobile RF communications equipment should be used no closer to any part of the
device, including cables, than the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with this symbol:
Compliance Level
± 6 kV contact
± 8kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
(> 95% dip
<5% U
T
T
) for 0.5 cycles
for 0.5 cycles
(60% dip in
40% U
T
T
5 cycles
70% U
T
T
) for
25 cycles
T
< 5%U
T
T
) for
for 5 sec
T
3 A/m
26 Specifications and Standards Compliance
Electromagnetic Environment
Guidance
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial and/or hospital
environment
Mains power quality should be that of a
typical commercial and/or hospital
environment
(> 95% dip in U
)
Mains power quality should be that of a
T
typical commercial and/or hospital
environment. If the user of the device
(60% dip in U
) for
T
requires continued operation during power
mains interruptions, it is recommended that
(30% dip in U
) for
T
the device is powered from an
uninterruptible power supply.
(> 95% dip in U
)
T
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial and/or hospital
environment
209