EDAN SE-1 User Manual page 63

Electrocardiograph
Table of Contents

Advertisement

SE-1 Electrocardiograph User Manual
Appendix 2 EMC Information
Guidance and manufacture's declaration – electromagnetic emissions-
Guidance and manufacture's declaration – electromagnetic emission
The SE-1 Electrocardiograph is intended for use in the electromagnetic environment specified
below. The customer or the user of the SE-1 Electrocardiograph should assure that it is used in
such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration – electromagnetic immunity –
Guidance and manufacture's declaration – electromagnetic immunity
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of Electrocardiograph should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
for all EQUIPMENT and SYSTEMS
Compliance
Group 1
Class A
Class A
Complies
For all EQUIPMENT and SYSTEMS
IEC 60601 test
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±1 kV for power
supply lines
supply lines
Electromagnetic environment – guidance
The SE-1 Electrocardiograph uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The SE-1 Electrocardiograph is suitable for
use in all establishments, other than domestic
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance level
- 56 -
EMC Information
Electromagnetic environment -
guidance
It is recommended the use of
antistatic materials. If floor are
covered with synthetic material,
the relative humidity should be at
least 50%.
It is recommended the use of
filters on power input lines and
enough separation between signal
lines and power lines.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents