Calibration Voltage:
DC Offset Voltage:
Minimum Amplitude:
Noise:
Filter
CMRR
Sampling Frequency
External Input/Output (Optional)
Input
Output
Note: Operation of the equipment below the minimum amplitude may cause inaccurate
results.
1mV±3%
±300mV
20 μVp-p
≤12.5µVp-p
EMG Filter: 35Hz (-3dB)
AC/DFT Filter: 50Hz/60Hz (-20dB)
≥110dB
1000 Hz
≥100kΩ; Sensitivity 10mm/V±5%;
Single ended
≤100Ω; Sensitivity 1V/mV±5%;
Single ended