Axis AKA/G Series User Manual page 45

These balances are destined for high accuracy weighing in laboratory practice.
Table of Contents

Advertisement

____________________________________________________________________________________________________________________
APPLICATIONS
...
Stats
...
Exit
STATS
1. Activation
2. Preview
3. Quantity:
4. Nominal value:
5. Tolerance:
6. Printout
7. Mode:
8. Taring
9. Shortcut:
Max ...
Min ...
0%
STA/PREVIEW
Quantity
In tolerance =
Under tol.
Over tol.
Total
Average
- print
F1-hist
USER MANUAL
10
5.0000 g
10%
<on><off>
<Auto><Manual>
<on><off>
F5
e= ...
d= ..
AUT TOT
2.0002
STATS
N = 2
=
10
5
=
3
=
3
= 50.000 g
= 5.0012 g
F2 – reset
CLR - exit
Applications option:
- Activation – move to weighing with summing,
- Preview – checking summing register state,
- Mode – add next result,
<Auto> - automatic after stabilization,
<Manual> - after putting load and pressing
key,
- Taring – summing with taring after each
weighing (without putting off the weight from
pan),
- Shortcut - fast access key selection: F1,
F2,... or F5.
I
Actions order:
Make a series of measurements using
after each measurement. In Auto mode
measurements are saved automatically.
g
100%
Each measurement inscribing is confirmed by
displaying sum and average value.
Choosing app and Preview option (or using
shortcut key) displays statistical results and
available options:
- stats register printout,
F1 – displaying histogram,
F2 – reset (zeroing) memory ,
CLR - back to summing.
In order to end Stats app choose the application,
choose Total and Deactivation.