Checkweighing Function (Thr) - Axis AKA/G Series User Manual

These balances are destined for high accuracy weighing in laboratory practice.
Table of Contents

Advertisement

____________________________________________________________________________________________________________________

16.10 Checkweighing function (thr)

This function allows comparing weighing result with two programmed reference values: lower and upper
threshold. Comparison result is signalled with indicators (MIN, OK, MAX) and sound signal generated (in
Impulse mode)when threshold values are exceeded.
-
Standard scale is set for cooperation with optical indicator.
On outputs P1-P3 (Relays socket) short-circuit states appear as result of comparison scale indication with
threshold values.
On the chart below output states are shown during increasing load on the scale for both working modes:
Indication mode:
P1
thr I
zero
treshold
P2
thr I
P3
In Batching mode on P1 (thr I) and P2 (thr II) outputs short-circuit impulses appears for time of 0,5s. On P3
(zero) output short-circuit state appears when indication does not exceed threshold value signalling zero
load.
USER MANUAL
Batching mode:
thr II
thr II
P1
thr I
P2
P3
zero
treshold
41
thr II