STATIM 900J Operator's Manual

1. Introduction

Figure 1
The STATIM
900J autoclave has been designed expressly for the rapid
®
sterilization of instruments intended for immediate use. This allows you to
sterilize instruments between patients. To decrease instrument turn-around
time, the Statim 900J does not have a post-sterilization drying cycle. Quick
drying and cooling of the load at the end of the cycle is achieved through
evaporation after the automatic opening of the sterilization chamber drawer.
Because of the unique operation of the Statim 900J, once the instruments are
exposed to ambient conditions, their continued sterility cannot be assured. It is
recommended to use these instruments immediately after they are comfortable
to the touch. The Statim 900J has three sterilization cycles:
The Solid Instrument Cycle is used to sterilize solid metal instruments. The
sterilization phase of this cycle is at 134°C for 3.5 minutes. For a detailed
description of this cycle, please see page 14.
The Hollow or Wrapped Instrument Cycle is used to sterilize dental handpieces.
The sterilization phase of this cycle is at 134°C for 5 minutes. For a detailed
description of the cycle please see page 15.
The Rubber or Plastic Instrument Cycle is used to gently sterilize rubber and
plastic instruments. The types of materials appropriate for this cycle are outlined
in Preparing and Loading Instruments. The sterilization phase of this cycle is at
121°C for 15 minutes. For a detailed description of this cycle please see page
15.