Installing the sensor

Prerequisites
1. 2 inch Tri-Clamp
2. 2 inch type 1 gasket
3. 2 inch tank ferrule or tee
About this task
The sensor may be installed in either a tank or pipe using a customer-supplied Tri-Clamp
and tee assembly. Keep at least 1 in. (25 cm) between the sensor and the pipe wall. If
clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely
submerged in liquid.
Figure 2-1: Rosemount 225 dimensional drawing
Procedure
1. Mount the sensor in the pipe.
2. Mount the sensor in a vertical pipe run with flow from top to bottom.
3. Ensure that the sensor is completely submerged in liquid to the flange.
Rosemount 225 Sensors
If the sensor must be mounted in a horizontal pipe run, orient the sensor in the
3 o'clock or 9 o'clock position.