3. OPERATING INSTRUCTION
(1)Make sure the ba=ery charger output voltage is the same as the connec6ng ba=ery.
(2)Plug in the power cord. LED indicates green flash when AC power on.
(3)Connect the ba=ery charger to the ba=ery. Start charging; please refer to LED INDICATION
4. LED INDICATION
(1)Green Flash:Power on
(2)Orange:Charging
(3)Orange Flash:Pre charge
(4)Green & Orange Flash:Charged 80%。
(5)Green:Full charged(Floa6ng charge)。
(6)Red Flash:Defect
5. TROUBLE SHOOTING
(1) If green indicator is off:
․Check AC input. If it works func6onally, the ba=ery charger may be defec6ve.
(2) If green indicator keeps flashing and cannot turn to charging indica6on:
․Check if the ba=ery connector is connected successfully.
․Check if there is any short circuit on the output connec6on.
․The ba=ery charger may be defec6ve if the ba=ery connec6on works func6onally.
(3) If red indicator keeps flashing:
․Check if the ba=ery connec6on is reversed.
․Check if there is any short circuit on the output connec6on.
․Check if the environment temperature is too low (0
․The ba=ery charger may be defec6ve if the red indicator s6ll keeps flashing.
(4) Charging indicator (orange) cannot turn to green:
․The ba=ery might be defec6ve, please stop charging and have the ba=ery be repaired.
(5) If the charging indicator (orange) turns to green (fully charged) immediately:
․The ba=ery may be in well-charged condi6on.
․The ba=ery may be defec6ve if the ba=ery is not fully charged.
6. CAUTION
(1) Before using the ba=ery charger, read all instruc6ons and cau6onary markings.
(2) Use the ba=ery charger in a well-ven6lated area
(3) To avoid the risk of injury, charge only lead-acid or gel cell type rechargeable ba=eries.