Gossen MetraWatt PROFITEST INTRO Operating Instructions Manual page 2

Din vde 0100-600 / iec 60364-6
Table of Contents

Advertisement

LCD Panel
Guide for
Carrying Strap
Fixed Function Keys
ESC: Return from
submenu /
Activate instrument
from standby state
MEM: Key for
memory functions
HELP: Access context
sensitive help
START:Switch on /
start measurement
I
:
Triggering key /
N
offset)
compensation (
LED Indications (see also section 16)
LED Case A
LED Case B
Mains
Mains
Green
Blinks green
Mains
Mains
Orange
Blinks red
LIMIT
Red
2
Control Panel
LCD Case A LCD Case B Function – Cause
A
Lights up green: Correct connection, measurement enabled
B
Blinks green:
PE
PE
x
L
N
L
N
A
Lights up orange: 2 different phases active (no neutral conductor at mains), meas. enabled
PE
B
Blinks red:
PE
C
Lights up red:
L
N
PE
L
N
x
L
N
UI, UIN:
IN
:
IF
:
After safety shutdown
RLO:
RINS, RE(INS):
Setup Menu
Softkeys
Parameter selection
Limit value specification
Entry functions
Memory functions
Neutral conductor not connected, measurement enabled
I
, I
, Z
-
, Z
-
, R
: No line voltage or PE interrupted
N
F
L
PE
L
N
E
RINS and RLO: Interference voltage detected, measurement disabled
Touch voltage > 25 or 50 V, measurement disabled: display: U.PE > UL!
During the tripping test with I
, the RCD is not tripped within 400 ms
N
(1000 ms for selective RCCBs of type RCD S).
With rising residual current, the RCD is not tripped before reaching I
The permissible (selected) limit value has been exceeded.
The permissible (selected) limit value has been exceeded.
Rotary
Selector Switch
Guide for
Carrying Strap
MAINS/NETZ LED
 see below
Limit LED
 see below
.
N
GMC-I Messtechnik GmbH

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROFITEST INTRO and is the answer not in the manual?

Table of Contents