Power Express system can include up to four Centrifuge modules
Configured for a single tube size (13x75 mm, 13x100 mm, or 16x100 mm) and
discipline (Chem/IA, Coag, or both)
Two types of centrifuges: ambient or temperature controlled
Accepts up to 40 tubes per load
Uses transfer arm with 5-grippers to transfer sample tubes from carriers on the track
into one of four sample holders within the centrifuge
Uses water-filled balance tubes, as needed, to create a balanced centrifuge load;
centrifuge will automatically turn off if there is a 20 gram load imbalance
Can be operated in Auto or Manual mode
Auto: Use for normal processing to load tubes automatically from Power Express
Manual: Use in error recovery to load sample tubes directly at the centrifuge
Centrifuge default spin time and speed is 4 minutes at 3000 rpm; adjustable at
centrifuge Control Panel
Temperature controlled centrifuge
Temperature control panel located on the front panel of the module; activates an
audible alarm and flashing beacon when temperature is outside of acceptable
range for more than the preset timer value (default is 30 minutes)
Maintains centrifuge bowl set point temperature of 21ºC +/- 3ºC
Loads sample tubes into the centrifuge until it is full or until load timeout expires
before another sample tube arrives at the module; use Priority button on D-Lane
Control Panel to override load timeout and start centrifugation immediately
Power Express In-Lab Training Manual
Version 2.0 (July 2016)
Centrifuge Module
Page 16 of 147