Installation Guide; Selecting An Installation Site - YOKOGAWA EXAxt ZR22G Technical Information

Direct in situ zirconia oxygen analyzer (applications volume)
Hide thumbs Also See for EXAxt ZR22G:
Table of Contents

Advertisement

2.

Installation Guide

2.1

Selecting an Installation Site

Improper installation of the detector may cause inaccurate measurement or damage in short-
term service.
(1) Easy access for inspection and maintenance.
(2) Ambient temperature does not exceed 150°C, and the terminal box is not exposed to
radiant heat.
(3) A clean environment free from corrosive gases.
(4) Minimum vibration.
Typical good/bad installation examples are shown below for your reference.
Bad Example
Duct
All Right Reserved Copyright © 2004, Yokogawa Electric Corporation
Gas Flow
(Oriented obliquely against
the gas flow)
(Installed on a bend in the duct)
(Mounted above the horizontal)
(Subjected to condensation)
(A corrosive gas or flammable gas
is contained in the atmosphere)
<2. Installation Guide>
Good Example
Probe Protector
Dust Guard Protector
Instrument air
(As reference gas)
TI 11M12A01-01E
2-1
F0201.ai
Mar. 14, 2017-00