B.3 Electromagnetic Immunity
Guidance and manufacture's declaration – electromagnetic immunity
iM3 is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3 should assure that they are used in such an environment.
IEC/EN 60601 test
Immunity test
3 V
Conducted RF
150 kHz to 80 MHz
IEC/EN
61000-4-6
6Vrms
bands between
Radiated RF
0,15 MHz and 80
MHz
IEC/EN
61000-4-3
3 V/m
80 MHz to 2.7 GHz
See table 1
Compliance
level
3 V
rms
150 kHz to
80 MHz
c
in
ISM
6Vrms
ISM
between
0,15
and 80 MHz
3 V/m
80 MHz to
2.7 GHz
Comply with
table 1
Electromagnetic environment -
level
Portable and mobile RF
communications equipment should be
used no closer to any part of iM3,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
rms
Recommended separation distance
d
1
2 .
c
in
bands
MHz
d
1
2 .
d
2
3 .
d=6
communications equipment bands
(Portable RF communications
equipment (including peripherals such
as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the iM3, including cables specified by
the manufacturer).
Where P is the maximum output
power rating of the transmitter in watts
(W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
- 130 -
EMC Information
guidance
150KHz to 80MHz
P
80 MHz to 800 MHz
P
800 MHz to 2.7 GHz
P
/E at RF wireless