Nibp - EDAN iM3 User Manual

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
A.3 NIBP
Complies with IEC 80601-2-30: 2009.
EDAN Module
Technique
Mode
Measuring Interval in AUTO
Mode (unit: minutes)
Continuous
Measuring Type
Average measurement
☆Measuring Range
☆Adult Mode
☆Pediatric Mode
☆Neonatal Mode
☆Alarm Type
☆ Cuff
Pressure
Range
Pressure Resolution
☆Maximum Mean Error
☆Maximum Standard Deviation
Maximum Measuring Period
Adult/Pediatric
Neonate
Typical Measuring Period
Oscillometry
Manual, Auto, Continuous, Average
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
5 min, interval is 5 s
SYS, DIA, MAP, PR
Interval (unit: minutes)
times
SYS: 40 mmHg to 270 mmHg
DIA: 10 mmHg to 215 mmHg
MAP: 20 mmHg to 235 mmHg
SYS: 40 mmHg to 230 mmHg
DIA: 10 mmHg to 180 mmHg
MAP: 20 mmHg to 195 mmHg
SYS: 40 mmHg to 135 mmHg
DIA: 10 mmHg to 100 mmHg
MAP: 20 mmHg to 110 mmHg
SYS, DIA, MAP
0 mmHg to 300 mmHg
Measuring
1 mmHg
± 5 mmHg
8 mmHg
120 s
90 s
20 s to 35 s (depend on HR/motion disturbance)
- 119 -
Product Specification
1/2/3/4/5
3/5

Advertisement

Table of Contents
loading

Table of Contents