EDAN iM3 User Manual page 80

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
1
Do not use this device near flammable anesthetics. Not suitable for use in the
presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
2
Do not use this thermometer without first installing a new probe cover.
3
Do not reuse the disposable probe covers.
4
Use probe covers supplied by the manufacturer with this thermometer only. Use of
any other probe cover will result in erroneous temperature readings.
5
The thermometer and probe covers are Non-sterile. Do not use on abraded tissue.
6
To limit cross contamination, use Blue devices for Oral and Axillary temperature
taking only.
7
Use RED devices only for RECTAL temperatures.
8
Thoroughly dry all electrical contacts on both probe and thermometer after washing,
or device may fail to function properly.
9
For re-calibration, service or integrity checks, refer to a qualified Biomedical
Technician or return to the manufacturer.
10 Do not open the F3000 module. No user-serviceable parts inside. Opening of the
module may affect calibration and voids warranty.
11 Disposal of used probe covers must be performed in accordance with current
medical practices or local regulations regarding disposal of infectious, biological
medical waste.
12 Cleaning frequency and practices must be consistent with institutional policy for
cleaning of non-sterile devices.
13 The F3000 module is not intended for neonatal patients.
14 In monitoring mode of TEMP module, no physiological alarms are available.
15 Verification of the temperature module is necessary as frequently as dictated by your
Hospital Procedures Policy. When you need to calibrate the temperature
measurement, please contact the manufacturer.
NOTE:
1. The materials with which the patient or any other person can come into contact
conform with the standard of EN ISO 10993:2009
2. Even though this device has been designed to minimize the effects of electromagnetic
interference, it does generate radio frequency energy. If not used in accordance with
the instructions, the device could cause interference in other equipment operating
within its vicinity. If the device is causing interference, the following actions may be
taken in an attempt to correct the interference:
• Re-orient or re-locate the receiving device.
• Increase the separation between the devices.
WARNING
- 70 -
Monitoring TEMP

Advertisement

Table of Contents
loading

Table of Contents