Vital Signs Monitor User Manual
☆Pediatric Mode
☆Neonatal Mode
☆Alarm Type
Pressure Resolution
☆Maximum mean error
☆Maximum standard deviation
Maximum measuring period
Adult
Pediatric
Neonate
Overpressure protection
Adult/Pediatric
Neonate
Pre-inflation Pressure
Adult Mode
Pediatric Mode
Neonatal Mode
SYS: 40 mmHg– 230 mmHg
DIA: 20 mmHg– 160 mmHg
MAP: 26 mmHg– 183 mmHg
SYS: 40 mmHg – 130 mmHg
DIA: 20 mmHg– 100 mmHg
MAP: 26 mmHg – 110 mmHg
SYS, DIA, MAP
1 mmHg
± 5 mmHg
8 mmHg
130 s
90 s
75 s
<300 mmHg
<150 mmHg
120 mmHg, 140 mmHg, 150 mmHg, 160 mmHg, 180
mmHg, 200 mmHg, 220 mmHg, 240 mmHg, 260 mmHg,
280 mmHg
Default: 160 mmHg
80 mmHg, 100 mmHg, 120 mmHg, 140 mmHg, 150
mmHg, 160 mmHg, 180 mmHg, 200 mmHg, 220 mmHg,
250 mmHg
Default: 140 mmHg
60 mmHg, 70 mmHg, 80 mmHg, 90 mmHg, 100 mmHg,
120 mmHg, 140 mmHg
Default: 90 mmHg