EDAN iM3 User Manual page 46

Vital signs monitor
Hide thumbs Also See for iM3:
Table of Contents

Advertisement

Vital Signs Monitor User Manual
Message
NIBP Interference
NIBP Leak Test Error
T2A TEMP
Temp COMM STOP
Temp exceed limit
No TEMP SENSOR
Ambient
temp
too
high
Ambient temp too low
Warm-up error
Cause
Signal noise is too large or
pulse rate is not regular due
to the patient movement.
Fail to deflate normally
during the leak test, so
NIBP leak test cannot be
finished.
TEMP module failure or
communication failure.
The TEMP value is beyond
the range of 35.5 C ~ 42 C
TEMP
sensor
is
connected to the TEMP
module.
The Sensor temperature is
higher than +40 C
The Sensor temperature is
lower than +10 C
Malfunction in the warm-up
circuit
- 36 -
Alarm Level
Low
Low
High
Medium
not
Low
Low
Low
Medium
Alarm Information
Action Taken
Make sure that the
patient
under
monitoring
is
motionless.
Test again. If the
problem still exists,
contact your service
personnel.
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.
Put the sensor into
the sensor bracket,
take
it
out
and
measure again.
Connect the sensor
and
the
monitor
well, and measure
again.
Put the sensor into
the sensor bracket,
measure again after
the
ambient
temperature reaches
normal value.
Stop
using
measuring function
of TEMP module;
notify
biomedical
engineer
or
Manufacturer's
service staff.