Vital Signs Monitor User Manual
33
The monitor should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, you must check that normal operation is
possible in the necessary configuration before you start monitoring patients.
34
Do not touch accessible parts of medical or non-medical electrical equipment in the
patient environment and the patient simultaneously, such as USB connector, VGA
connector or other signal input/output connectors.
35
SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
36
SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
37
Operation of the equipment exceeding specified physiological signal or the
operational specification may cause inaccurate results.
38
The equipment can provide protective means to prevent the patient from being
burned when used with HF SURGICAL EQUIPMENT. The equipment can protect
against the effects of the discharge of a defibrillator. Use only EDAN-approved
accessories.
39
When the monitor is used with HF surgical equipment, the transducer and the cables
must be avoided from conductive connection to the HF equipment. This is to protect
against burns to the patient.
40
To protect the monitor from damage during defibrillation, for accurate measurement
information and to protect against noise and other interference, use only accessories
specified by EDAN.
41
No modification of this equipment is allowed without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe operation.
42
Clinical decision making based on the output of the device is left to the discretion of
the provider.
43
The monitor is equipped with Wi-Fi to receive RF electromagnetic energy. Therefore,
any other equipment complying with CISPR radiation requirements may also interfere
with the wireless communication and make it interrupted.
44
If the earth protection system is not stable, use the batteries for power supply.
WARNING