the PATIENT or other leads is avoided. Temporary unused ACTIVE ELECTRODES should be
stored so that they are isolated from the PATIENT.
10) For surgical procedures where the h.f. current flow through parts of the body having a relatively
small cross-sectional area, the use of bi-polar techniques may be desirable in order to avoid
unwanted coagulation.
11) Apparent low output or failure of the HF SURGICAL EQUIPMENT to function correctly at the
normal operating settings may indicate faulty application of the NEUTRAL ELECTRODE or poor
contact in its connections. In this case the application of the NEUTRAL ELECTRODE and its
connections should be checked before selecting higher power.
12) The supplier will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the USER's appropriately qualified
technical personnel to repair those parts of EQUIPMENT which are designed by the
manufacturer as repairable.
13) It may cause a malfunction when the unit is used in an environment that has strong magnetic
fields i.e. MRI room.
14) Do not use the unit other than your intended purpose. It may cause unexpected side effects.
15)Attention is called to the danger of ignition of endogenous gases
Sales &
Address:
Service Center
Tel :
Fax :
Website:
Address:
Tel:
3, Yeoksam-ro 33-gil, Gangnam-gu, Seoul, Korea
+82-2-558-1711∼6
+82-2-554-0317
www.daiwha.com
mdi Europa GmbHLangenhagener Str. 71
D-30855 Hannover-Langenhagen, Germany
+49-511-3908 9530
1. Precautions9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Meditom DT-400S and is the answer not in the manual?

Related Products for Daiwha Meditom DT-400S

This manual is also suitable for:

Meditom dt-300s

Table of Contents

Save PDF