Output Setting Guide - Daiwha Meditom DT-400S User Manual

Table of Contents

Advertisement

Part
Cut Mode
Pure
Skin Incision
or Blend1
Muscle
Pure
Dissection
or Blend1
Tumor
Blend3
Excision
Stomach,
Blend2
Intestine
or Blend3
Resection
Hemostasis
Massive
Coagulation
Neuro Surgery
Blend2
or Blend3
Prostatic
Pure
Resection
or Blend1
Bladder
Fulguration
Blend2
Tumor Excision
Blend3
Cervical
Blend2
Conization
Bartholin
Blend3
and Skeneis
Tubal
Blend2
Proctologic
Blend3
Abscess/Cyst
Blend3
Rectal, Sigmoid Blend3
Endoscopic
Pure
surgery

7. Output Setting Guide

Cut Power
Coag. Mode
10~120
8~100
Above 15
Above 15
Above 10
20 and up
20 and up
Contact
Spray
Loop
20 ~ 80
Contact
20 ~ 70
65 and up
Spray
55 and up
Spray
15 ~ 80
Spray
15 ~ 70
10 ~ 80
Contact
15 ~ 30
Spray
8 ~ 50
8 ~ 40
Spray
10 ~ 80
Spray
Lancet 8~30
Contact
Loop 10~20
25-40
Spray
Coag. Power
BI-Coag.
10~65
3~22
10~30
BALL
1~8
10~25
15 and up
12~30
10~15
25~70
12~22
10~20
10~22
10~15
12~30
30-50
7. Output Setting Guide37

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Meditom DT-400S and is the answer not in the manual?

Related Products for Daiwha Meditom DT-400S

This manual is also suitable for:

Meditom dt-300s

Table of Contents

Save PDF