3.1 Specifications

Product Name
Electrosurgical Equipment
Model
DT-400S/300S
Purpose
Cutting or coagulation of biological tissue using high-frequency current
Mains and Line
120(60Hz) or 230VAC(±10%) (50 or 60Hz)/switchable
Frequency
See the label on rear side for correct information
Two of F5.0A, 250V for 230VAC
Fuses
Two of F10.0A, 250V for 115 VAC
Power Consumption
900VA
Protection Class
Class 1, Type CF
IPX
Foot Switch: IPX6
Carrier Frequency
400 KHz ±10%, 482KHz ±10%
Repeat Frequency
33KHz
Low Freq. Leakage
EN60601-1 [2012] Requirement
Current
High Freq. Leakage
Less than 150mA
Current
Operation Conditions 10℃~40℃, 30%~ 85% noncondensing, 700mbar~1060mbar
Storage Conditions
-10℃~60℃, 30%~ 95% noncondensing, 700mbar~1060mbar
-10℃~60℃, 30%~ 95% noncondensing, 700mbar~1060mbar
Transportation
Do not drop the unit from higher than 50cm
Conditions
Do not use the hook.
Installation
5cm clearance required on each side unit for cooling
Conditions
Cooling
1 fan mounted
Recommended Duty
10 sec ON, 30 sec, OFF(Output for 10 seconds, pause for 30 seconds )
Cycle
Size (H, W, D)
147mmx 330mm x 417mm
Weight
About 16Kg
Not AP/APG device
APG
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with or with oxygen or nitro oxide
Operation Sound
Difference in operation sounds between cutting and coagulation.
Operation Display
Visible difference in operations between cutting and coagulation.
Output Setting
Displayed in [W] unit from Seven Segment (LED number displayer)
changed without notice
It can be
to improve performance.
3. Specifications
3. Specifications19

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Meditom DT-400S and is the answer not in the manual?

Related Products for Daiwha Meditom DT-400S

This manual is also suitable for:

Meditom dt-300s

Table of Contents

Save PDF