Daiwha Meditom DT-400S User Manual page 31

Table of Contents

Advertisement

3) Select Single Pad Mode or Dual Pad Mode
▶ Single Pad Mode
① Connect the Single Pad to the Return Electrode Jack. The Alarm will stop and it will be
ready to be used.
② Attach the Pad firmly to the patient. (Please read carefully the precautions)
▶ Dual Pad Mode
① Connect the Dual Pad to the Return Electrode Jack.
② Press
(Single Pad or Dual Pad Selection Switch) to set up the Dual Pad Mode.
The Dual Pad Mode Indicator will be turned on and the DRM will be ready.
③ Attach the Pad firmly to the patient. (Please read carefully the precautions)
④ The changes in the Assist Resistance Indicatorwill be seen. You can set DRM by pressing
(DRM Setting Switch) if the LED light on from 3 to 8 steps.
※ The Assist Resistance Indicator has 2 steps or 0 step when the Return Electrode cable
may be cut off or the Pad is not attached properly to the patient. The Indicator has 8 steps
when the Return Electrode cable may be short-circuited.
6. Operating and Sequence31