5) Patch Neutral Plate as closely and firmly to the surgery area as possible and secure as the large
contact area as possible.
6) Method and sequence to operate each of functions
6.1) When operating Mono-polar cut function
-Connect TBH or Uni-polar Handle to 'ono-polar RF Power Outlet'.
- Press 'Cutting Mode Selection Switch' to select the proper output mode for the surgery. The
indicator is on for the selected output mode, and the value of applicable mode will be
displayed in the 'Cutting RF Power Digital Display'.
- Use the Cutting RF Power Increase/Decrease Switch to configure the cutting output.
- Press the yellow switch of TBH, or the yellow pedal of Foot Switch to perform the cutting
surgery.
6.2) When operating Mono-polar coagulation function
-Connect TBH or Uni-polar Handle to Mono-polar output jack.
- Press 'Coagulation Mode Selection Switch' to select the proper output mode for the surgery.
The indicator is on for the selected output mode, and the value of applicable mode will be
displayed in the 'Coagulation RF Power Digital Display'.
- Use the 'Coagulation RF Power Increase/Decrease Switch' to configure the coagulation output.
- Press the blue switch of TBH, or the blue pedal of Foot Switch to perform the coagulation
surgery.
6.3) When operating bi-polar cut function
- Connect bi-polar forceps to 'Bi-polar RF Power Outlet'.
- Press 'Cutting Mode Selection Switch' to select Bi-Cut mode for the surgery. The indicator is
on for the selected output mode, and the value of applicable mode will be displayed in the
'Cutting RF Power Digital Display'.
- Use the 'Cutting RF Power Increase/Decrease Switch' to configure the cutting output.
- Press the yellow pedal of Foot Switch to perform the cutting surgery.
6.4) When operating bi-polar coagulation function
- Connect bi-polar forceps to 'Bi-polar RF Power Outlet'.
- Press 'Coagulation mode Selection Switch' to select Bi-Coag. mode for the surgery. The
indicator is on for the selected output mode, and the value of applicable mode will be displayed
in the 'Coagulation RF Power Digital Display '.
- Use the 'Coagulation RF Power Increase/Decrease Switch' to configure the coagulation output.
- Press the blue pedal of Foot Switch to perform the coagulation surgery.

6.3.3 Guidelines after use

1)When the operation of ESU has been finished, turn off power and disconnect all of accessories.
2) Accessories should be kept in the clean environment, and patient contacting components should
be sterilized.
3) Please sterilize reusable accessories before you keep in store.
6. Operating and Sequence35

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Meditom DT-400S and is the answer not in the manual?

Subscribe to Our Youtube Channel

Related Products for Daiwha Meditom DT-400S

This manual is also suitable for:

Meditom dt-300s

Table of Contents

Save PDF