Operating And Sequence; Monopolar - Daiwha Meditom DT-400S User Manual

Table of Contents

Advertisement

6.1 Monopolar

1) After checking the main power voltage and frequency, the power cord should be connected to
the main power inlet located at the rear panel of the equipment. If there is no ground circuit in
the main power outlet, the Equipotential grounding cable must be connected to the quipotential
Grounding. Contact Daiwha sales department when you need the Equipotential Grounding
cable.
2) The Power Switch is turned on - If the Pad is not connected, The Return Fault Indicator will be
twinkle and the alarm will ring to warn the fault of the Return Electrode.

6. Operating and Sequence

6. Operating and Sequence30