BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _
Technical Description
english
User's Guide
http://www.balluff.de
http://www.balluff.de

Advertisement

Table of Contents
loading

Summary of Contents for Balluff BTL5-E Series

  • Page 1 BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _ Technical Description english User's Guide http://www.balluff.de http://www.balluff.de...
  • Page 2: Table Of Contents

    DATech for Testing Electro- Fast transients (Burst) magnetic Compatibility, has con- EN 61000-4-4 Severity level 3 firmed that Balluff products meet The following patents have been Surge the EMC requirements of the fol- EN 61000-4-5 Severity level 2...
  • Page 3: Function And Characteristics

    BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _ Micropulse Linear Transducer in Profile Housing Function and Characteristics Characteristics the damping zone. The wave arriv- Available stroke lengths ing at the beginning of the and magnets Micropulse transducers feature: waveguide creates an electrical To provide for optimum fit in any —...
  • Page 4: Floating Magnets

    BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _ Micropulse Linear Transducer in Profile Housing Installation (cont.) Nominal stroke Mounting brackets with isolation bushings and M5 x 22 cylinder head screws, DIN 912, max. tightening torque 2 Nm Fig. 3-1: Dimensional drawing (BTL5...P-S 32 transducer with floating magnet BTL5-P-3800-2) max.
  • Page 5: Captive Magnets

    BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _ Micropulse Linear Transducer in Profile Housing Installation (cont.) Nominal stroke Black, round marking Magnet Mounting brackets with isolation bushings and M5 x 22 cylinder head screws, DIN 912, max. tightening torque 2 Nm Fig.
  • Page 6: Wiring

    To ensure electromagnetic com- ➃ limited protection. patibility (EMC), which Balluff Reference GND for supply verifies by the CE Marking, the voltage and EMC! Cable length max. 20 m; Ø 6 to following points must be strictly 8 mm.
  • Page 7: Startup

    BTL5-A/C/E/G_ _-M/U_ _ _ _-P-S 32/KA_ _ Micropulse Linear Transducer in Profile Housing Wiring (cont.) 4...20 mA 10...0 V 0...10 V +24 V Fig. 4-3: BTL5-E10...KA _ _ with processor card/controller, wiring example Startup Check connections Voltage output 10 V ponents should be regularly increasing decreasing...
  • Page 8: Technical Data

    ➥ ➥ ➥ ➥ ➥ Fig. 3-7 Output current 4...20/20...4 mA BTL5-N-2814-1S Individual specifications as per Weight approx. 35 g Load resistance < 500 Ohm Balluff factory standard Housing anodized aluminum Contact surface plastic Dimensions, weights, Connection to processor ambient conditions Control arm (optional)