Step 4
Compare control solution result to the range
printed on the TEST STRIP VIAL.
Example
If the results are not within the control range printed on the test strip
vial, the meter and test strips may not be working properly. Repeat
the control solution test.
18
CONTROL
RANGE
mCAUTION: The
control solution range
printed on the test strip
vial is for OneTouch
Ultra
Control Solution
®
only. It is not a
recommended range for
your blood glucose level.
®