Procedure
1. Enter the Main menu, first by pressing any key to get the enter main menu
2. Select the up arrow key (menu button 1 or reed switch 1) or down arrow key
3. Select the Enter key (menu button 3 or reed switch 3) to enter the option.
4. If configuring channel 1's gas type, select it with the Enter key (menu button 3 or
5. If configuring channel 2's gas type, use the down arrow key (menu button 2 or reed
6. If the gas type displayed is not the required target gas, select the down arrow key
7. Activate the enter key (menu button 3 or reed switch 3) to select the specific gas
8. Use the down arrow key (menu button 2 or reed switch 2) and then the Enter key
9. Enter the main menu, first by pressing any key to get the enter main menu
10. When Calibrate Sensor? is displayed, select the Enter key (menu button 3or
11. When Calibrate Sensor #1? is highlighted, press the Enter key (menu button
12. If calibrating sensor 2, select the down arrow key (menu button 2or reed switch 2)
13. When the desired sensor to be calibrated (1 or 2) is highlighted, activate the enter
14. Select YES (menu button 1or reed switch 1) to confirm the selection.
15. Apply clean air when Apply Clean Air is displayed; then select Z & Span using
Net Safety SC311
prompt, then press menu button 1 (reed switch 1) to select yes .
(menu button 2 or reed switch 2), until Select Gas Type is displayed.
reed switch 3.
switch 2) and then select it with the Enter key (menu button 3 or reed switch 3).
(menu button 2 or reed switch 2 until the desired target gas is found.
to be detected.
(menu button 3or reed switch 3) until the main menu is completely exited.
Refer to the Net Safety Millennium II Transmitter calibration procedure below
and/or
Figure 5-2
before attempting calibration. If the sensor's target gas is setup
correctly as desired, follow the steps below for full calibration/normal calibration
procedure.
Important
Always use 50 percent span gas (half the scale) of the specific target gas when
calibrating this sensor.
Note that if a calibration is not successful the transmitter will display the message
, and you will have to initiate a manual reset. Refer to the Net Safety
Span failed
Millennium II Multi-Channel Transmitter
prompt, and then pressing or selecting menu button 1 or reed switch 1 to select
.
yes
reed switch 3).
3or reed switch 3) if calibrating this sensor.
to scroll to Calibrate Sensor #2 ?
key (menu button 3or reed switch 3).
( (menu button 1or reed switch 1) for normal calibration.
The transmitter will display Setting zero as the sensor is being zeroed. (Ensure
no contaminant gases are around if using ambient air).
Reference Manual
for manual reset.
Maintenance
May 2019
21