Configure And Install The Supplicant Configuration File - Cardiac Science Burdick 8300 Network Setup Manual

Electrocardiograph
Hide thumbs Also See for Burdick 8300:
Table of Contents

Advertisement

Supplicant Configuration Details

Configure and install the supplicant configuration file

Caution: Possible data corruption
This procedure requires you to access the Service Functions menu. The Service Functions
menu is intended for qualified Service personnel. If you have any question, please contact
Technical Support.
To configure a supplicant configuration complete the following steps:
Install the configuration file on the ECG device:
1.
Insert the USB device containing the supplicant configuration file into the USB
a.
port on the back of the ECG device.
Select Proceed to Service functions.
b.
Press Enter.
c.
Use the navigation keys to select Import wpa supplicant config and then press
d.
Enter.
The system copies the file and displays a message indicating the transfer was
successful.
Install the digital certificate on the ECG device (optional).
2.
Export the required certificate to a file on a PC.
a.
Use the pem format.
Note: Your config file must contain lines that specify the name and location of your
digital certificate. Certificates are located at this path after they are copied to the ECG
device.
/atlas/DATA/pdofiles/wpa_supplicant_config directory.
Copy the certificate file to a blank USB device.
b.
Insert the USB device containing the certificate file into the USB port on the back
c.
of the ECG device.
Press Setup.
d.
Select Service functions.
e.
The ECG device displays a warning message.
Press Enter.
f.
Use the navigation keys to select Import wpa supplicant certificate and then press
g.
Enter.
The system copies the file and displays a message indicating the transfer was
successful.
70-00900-02 C
A-2
Burdick 8300/8500