Troubleshooting; Specifications - PCE Health and Fitness PCE-PWT 10 User Manual

Salt meter
Table of Contents

Advertisement

19

Troubleshooting

LCD
CAUSE
DISPLAY
Electrode dried out
Measured value is out of
range
Incorrect calibration solutions
Setting value does not match
calibration solution
Electrode is broken
20

Specifications

Salinity
Conductivity
Temperature
Others
CORRECTIVE ACTION
Soak the conductivity electrode in tap water for 10
minutes
Check the electrode whether clogged, dirty or
broken
Using the fresh conductivity standard solutions for
calibration
Reset the calibration value
Replace the electrode module
Model
Range
Accuracy
Resolution
Range
Accuracy
Resolution
Calibration Points
Calibration Solutions
Range
Accuracy
Resolution
Calibration Points
Calibration Range
Temperature Compensation
Temperature Coefficient
Cell Constant
Normalization Temperature
Hold Function
Power Off
Sensor Type
Power Requirements
Dimensions
Weight
© PCE Instruments
13
PCE-PWT 10
0.00~10.00ppt
± 1% F.S
0.01ppt
0~20.00,
200.0,
2000µS/cm,
20.00mS/cm
± 1% F.S
0.01, 0.1, 1
1 to 3 points
10µS/cm,
1413µS/cm, 12.88mS/cm
0~60°C, 32~140°F
± 1°C
0.1°C, 0.1°F
1 point
Measured value ± 10°C
0~60°C, 32~140°F
2%/°C
K=1
25°C, 77°F
Manual or Automatic
Manual or Automatic (8 minutes
after last key pressed)
Order Code: ECSCAN-C1-10K
2× 1.5V "AAA" Batteries
185(L)× 40(Dia.)mm
100g
84µS/cm,