11
Troubleshooting
11.1
OR-UR / dilution
OR = Overrange / UR = Underrange
The test result is outside the measurement range of this method. OR results can be brought into
the measurement range by dilution. Use the dispensing pipette to take a 5 ml (or 1 ml) sample.
Fill the sample into the cuvette and add 5 ml (9 ml) of distilled water. Perform the measurement
and multiply the result by 2 (or 10). Dilution is not applicable to the parameter "pH".
11.2
Error codes
Error code
BAT!
Err02
Err03
Err04
Err05
12
Cuvette replacement
1.
Before replacing the cuvette, make sure that the instrument is dry and clean.
2.
Remove the old cuvette and dispose of it appropriately.
3.
Make sure that the new cuvette is clean.
4.
Insert the new cuvette and turn it until it locks into the holder. This may require some
force.
5.
To calibrate the instrument to the new cuvette, follow the procedure in chapter 9.1.3
Calibrat.
Descripton
Replace batteries
(too dark) Clean measuring chamber and dilute water sample
(too bright) Do not forget the light protection cover during the
measurement
Repeat ZERO and TEST procedure
Environmental temperature below 5 °C or above 60 °C
© PCE Instruments
36