11
3 Point Calibration
1.
If you have selected 3 points calibration in the setup menu, when second calibration
point is done, the display will immediately show "CAL3". The tester prompts you to
continue with third point calibration.
2.
Repeat the above steps until the display shows "END", the tester returns to
measurement mode. Calibration is completed.
Performing the conductivity calibration will simultaneously calibrate the corresponding
salinity value.
12

Temperature Calibration

During the measurement, if the temperature reading displayed differs from that of an accurate
thermometer, you need to calibrate the tester.
Press and hold the CAL key for 3 seconds to enter setup menu.
1.
Press ENTER key until tester shows the "
2.
Press CAL key to enter the temperature calibration mode.
3.
Press CAL key again to set temperature value (Resolution: 0.5
4.
Press ENTER key to confirm, the display shows next option.
5.
Press MEAS key, the tester returns to measurement mode. Calibration is completed.
6.
EXIT THE CALIBRATION:
During the calibration process, press MEAS key, the tester will return to measurement mode.
o
C/CAL" or "
Enter
© PCE Instruments
10
o
F/CAL".
o
C).