Chapter 13 Technical Data - Fisher Scientific Isotemp Series User Manual

Gravity and convection incubators
Table of Contents

Advertisement

Technical Data
The technical data are valid only for an empty device equipped with three shelves, a spray-painted outer enclosure
and a power line voltage 120 V/60 Hz. Options may have an impact on the specified performance.
Table 13-1 Technical Data - Gravity Convection Incubators
Parameter
Process
Work space atmosphere
Min.
Max.
Temperature deviation from set value at 37 °C
(98.6 °F), spatial. Max. value /typical value
Temperature deviation from set value at
37 °C (98.6 °F), over time
Heat-up time (work space unoccupied, from
25 °C (77 °F) to 98% of set temperature of
37 °C/98.6 °F)
Recovery time (work space unoccupied,
door open for 30 s, to set temperature).
Max. value/typical value
Heat dissipation to environment (at set
temperature of 37 °C (98.6 °F) and room
temperature of 25 °C/77 °F)
Overall dimensions
Height
Width
Depth
Overall weight
Loading capacity
Loading capacity per shelf
Max. overall loading capacity per device
Electrical data
Power consumption
Maximum current
Fisher Scientific
Isotemp 60L
Isotemp 100L
Unit
Incbtr Grvity
Ambient
°C/°F
temperature
plus 5 °C/41°F
°C/°F
75 °C/167 °F
K
± 0.7/±0.6
K
±0.2
min
52
min
5/3
W
21 ±10%
mm/in
720/
28.3
mm/in
530/
20.8
mm/in
565/
25.2
kg/lbs
40/88
kg/lbs
kg/lbs
50/110
W
300
A
2.5
Gravity Convection Incubators
13
Isotemp 180L
Incbtr Grvity
Incbtr Grvity
Ambient
Ambient
temperature
temperature
plus 5 °C/41°F
plus 5 °C/41°F
75 °C/167 °F
75 °C/167 °F
± 0.7/±0.6
± 0.7/±0.6
±0.2
±0.2
42
47
5/4
9/8
26 ±10%
31 ±10%
820/
920/
32.3
36.2
640/
640/
25.2
25.2
565/
738/
25.2
29.1
51/112
65/143
25/55
50/110
75/165
540
720
4.5
6.0
13-1