Safety Notes
Operational Safety Rules
Operational Safety Rules
The following rules must be heeded when working with Fisher Scientific incubators:
• Observe the sample weight limits specified for your Fisher Scientific incubator as a whole
and its shelving in particular; see
• Do not load the bottom of the interior workspace to avoid the risk of overheating any
samples placed there and to prevent the temperature sensor from being damaged.
• Arrange the samples evenly throughout the work space, making sure not to place them too
closely to the interior walls to ensure a uniform temperature distribution.
• Do not load your Fisher Scientific incubator with substances that exceed the capabilities of
the available lab apparatus and Personal Protection Equipment to provide sufficient
degrees of protection to users and third parties.
• Check the door seal every six months for proper sealing performance and possible damage.
• Do not process any samples containing hazardous chemical substances that may be
released into the ambient air through defective seals or may cause corrosion or other
defects on parts of the Fisher Scientific incubator.
Warranty
Fisher Scientific International, LLC warrants the operational safety and functions of the Fisher
Scientific incubators only under the condition that:
• the incubator is operated and serviced exclusively in accordance with its intended purpose
and as described in these operating instructions,
• the incubator is not modified,
• only original spare parts and accessories that have been approved by Fisher Scientifi care
used (third-party spares without Fisher Scientific International, LLC approval void the
limited warranty),
• inspections and maintenance are performed at the specified intervals,
• an operation verification test is performed after each repair activity.
The warranty is valid from the date of delivery of the incubator to the customer.
1-2
Gravity Convection Incubators
"Technical Data"
on
page
13-1.
Fisher Scientific