Warnings And Hazards - gbo Medizintechnik Duodynator Instructions For Use Manual

Table of Contents

Advertisement

Duodynator

7 Warnings and hazards

Warning
 For patients with implanted electronic device electrical stimulation
treatment is to be carried out only after having checked any risks.
 Turn off cellular phones and radiophones or place them in a distance
of 3 m from the device.
 Cardiac pacemakers are extremely vulnerable. Here the therapy should
only be carried out under continuous pulse and ECG control.
 If the patient and/or the patient cable is in the direct range of a high-
frequency, short-wave or micro-wave therapeutic device, a damage to
the device or an injury to the patient cannot be excluded. Please keep a
clearance distance of 3 m.
 A simultaneous connection of the patient to a high-frequency surgery
device can lead to burns under the electrical stimulus electrodes.
 The unit is not designed to be operated in places with the inherent risk
of explosions. If it is used in dangerous areas of anesthesia
departments, the possibility of an explosion cannot be excluded.
 Current densities exceeding 2 mA/cm² require the special attention
to the operator.
 It is recommended not to exceed the current density of 0,2 mA/cm
in case of galvanization and 0.1 mA/cm
 If currents with galvanic portions are used special attention have to
be kept. Keep care that the electrode sponges are always moistened
very good.
 Do not use self-adhesive electrodes at currents with galvanic
portion; they contain corrosive!
 The electrodes must be (fully covered) rest with the entire electrode
surface in order to avoid local current density peaks. Otherwise,
skin redness may occur in severe cases even burns.
Note!
!!
In
gbo Medizintechnik AG or one of the service agencies authorized by gbo
Medizintechnik AG immediately.
 gbo Medizintechnik AG
!
case
of
in case of iontophoresis.
2
any
visible
31
failure
contact
Version 1.2