3.2 Suction flyout

When using the Duodynator together with the Neoserv the suction operation will be
activated after the Neoserv is switched on. Pressing the suction button, a flyout
folded up in which you can set the parameters for the suction.
The suction, which means the negative pressure in the suction electrodes, can be
adjusted from -40 to -300 mbar.
The suction massage can be adjusted between 0 and 20 pulses / minute.
Note!
!
 If the suction is to be used, the Neoserv must be switched on with its
power switch on the back of the unit before.
 If the Duodynator is switched off using the power switch, the
Neoserv sounds a warning signal which indicates that the Neoserv is
still on and must be switched off.
Note!
!
 If the Neoserv is switched on and connected to the current
stimulation output of Duodynator, suction will automatically
switched on when selecting a current from the Favorite list.
 gbo Medizintechnik AG
Suction
Button
Duodynator
Suction strength
suction massage
Suction on-off
switch
Version 1.2