4.4.2 Procedure for combination therapy

Select the current form from the Duodynator.
1.
Set the desired modifications.
2.
White plugs has a positive polarity in the primary position, while the counter
3.
electrode - the metallic surface of transducer - is negative. Connect the white plug
of the patient cable to the neutral electrode and the black one with the Sonostat.
Apply the neutral electrode.
4.
Select the desired treatment head by pressing the button head left and head right.
5.
Choose an own therapy, an indication or the individual setting on Sonostat.
6.
Bring plenty of contact gel to the sonicated area of the patient. Move the
7.
transducer surface under light pressure to the skin.
Set the current intensity so that the patient experiences a significant irritant effect
8.
on the stimulation current therapy unit by turning the intensity knob. If the patient
feels a pain, reduce the stimulation current intensity.
Press the Start button to begin the Sonostat to the ultrasonic treatment now.
9.
Move the transducer surface under light pressure with circular or straight even
10.
strokes on the area to be sonicated.
At the end of treatment, the selected in the settings Therapy-end signal is heard.
11.
The power is automatically turned off and the ultrasonic intensity is automatically
set to 0.0.
Clean the probe from the contact means.
12.
When the treatment duration of the Duodynator is the elapsed, the selected in the
13.
settings Therapy-end signal and the intensity is automatically set to 0.0 sound.