Specifications - ECD H10C Instruction Manual

Microprocessor based conductivity, tds, salinity, temperature hand held meter
Table of Contents

Advertisement

Specifications

Display
Conductivity
Salinity
Temperature
Reference Temperature
Temperature Coefficient
Cell Constant
TDS Constant Range
Power
Calibration Back-up
Audio Feedback
Display
Ambient Temperature
Relative Humidity
Case
Dimensions (W x D x H)
Weight
9
Range
0.0 to 499.9uS/cm
500 to 4999uS/cm
5.00 to 49.99mS/cm
50.0 to 200.0mS/cm
0.0 to 70.0ppt
14°F to 194°F
(-10.0 to 90.0 °C)
59 to 77°F (15.0 to 25.0 °C)
0.0% to 4.0%
0.5 ± 0.15
0.30 to 1.00, default at 0.65
Four "AAA" Batteries
EEPROM
All Touch Keys
2" x 1 3 8 ⁄ " LCD
0 to 50 °C
up to 90%
IP67 water tight case
3.0" x 6.2" x 1.4" (75mm x 157mm x 35mm)
8.1 oz (230 g) (Batteries included)
Resolution
Accuracy
0.01uS/cm
±1% of reading + 2uS/cm
1uS/cm
±1% of reading + 5uS/cm
0.01mS/cm
±1% of reading + 0.05mS/cm
0.1mS/cm
±2.5% of reading + 0.5mS/cm
0.1ppt
±0.2% Full Scale
0.1 °C
±0.2°C or ±0.4% Full Scale, whichever is
greater.