Specifications

Display
pH
mV
Temperature
pH buffer recognition
pH Temperature compensation
pH Buffer Temperature range
pH Electrode Offset recognition
pH Electrode Slope recognition
Input Impedance
Temperature sensor
Power
Calibration Back-up
Audio Feedback
End Point Sensing & Hold
LCD Display
Ambient Temperature
Relative Humidity
Case
Dimensions
Weight
Range
0.00 to 14.00 pH
-1999.9 to 1999.9 mV
32 to 212°F (0.0°C to 100.0
°C)
Resolution
0.01 pH
0.1 mV
32.18° F (0.1 °C)
pH 7.00, 4.01, 10.01 or pH 6.86, 4.00, 9.18
AUTO/MAN 32 to 212°F (0.0°C to 100.0 °C)
0°C to 60.0°C
±100 mV at pH 7.00 or +108.3/-91.7mV at pH6.86
±30% at pH 4.00, 4.01, 9.18 and 10.01
12
>10
Thermistor, 10 kΩ at 77°F (25°C)
Six "AAA" Batteries
EEPROM
All Touch Keys
Yes
3 1 2 ⁄ " x 2 1 4 ⁄ "
32 to 122°F (0 to 50 °C)
up to 90%
IP54 watertight case
5.9" x 8" x 2.8" (150mm x 203mm x 72mm)
17.8 oz. (504 g) (Batteries included)
14
Accuracy
±0.1% pH
±0.1%/±0.1%
mV
±32.36°F (0.2 °C)