Remove the battery access cover.
3.
Slide the plate toward the front of the ECG device.
a.
Lift up the back end of the plate to remove.
b.
Remove battery pack:
4.
Slide the battery pack out of the battery compartment.
a.
Caution: Equipment damage.
Do not pull on the wires to disconnect the battery pack connector. This may cause the
connector to fail. Only pull on the connector to disconnect the battery pack.
Disconnect the battery pack connector.
b.
Remove the ferrite from the old battery pack by unsnapping fasteners.
c.
Figure 4-1: Example of ferrite removal for the Burdick 8300
Place the ferrite around the new battery pack cable approximately in the middle of
d.
the cable.
Burdick 8300/8500
Disassembly
70-00844-02 B
4-13