If you need to return the device to the manufacturer or supplier, please fill out the form • contained on the CD-ROM and send it with the device. Unfortunately, the manufacturer cannot repair or inspect the device without the completed form. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
• Remove the battery from device before returning to the manufacturer for service or warranty reasons. • Dispose battery packs in accordance with local regulations; where possible, recycle used batteries. 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
5 Signal cable (remote versions only) INFORMATION! Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
• Remote version (measuring sensor with connection box and a signal converter in a remote (field) housing) Figure 2-2: Device versions 1 Compact version in aluminium (IP67) housing 2 Compact version in polycarbonate (IP68) housing 3 Remote version 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
• Do not lift the device by the signal converter housing. • Do not use lifting chains. • To transport flange devices, use lifting straps. Wrap these around both process connections. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
• Allen key (4 mm) • Small screwdriver • Wrench for cable glands • Wrench for wall mounting bracket (remote version only) • Torque wrench for installing flowmeter in pipeline 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
WATERFLUX 3070 2.7.6 Magnetic field Figure 2-9: Avoid magnetic fields 2.7.7 Open discharge Figure 2-10: Installation in front of an open discharge 2.7.8 Bends Figure 2-11: Installation in bending pipes www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
Figure 2-12: Installation in bending pipes 2.7.9 Control valve Figure 2-13: Recommended installation: in front of a control valve 2.7.10 Air venting Figure 2-14: Air venting 1 ≥ 5 m 2 Air ventilation point 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
(IP66/67). Submersion under water is possible down to a depth of 5 meters. The remote IFC 070 signal converter is available in: • an aluminium housing suitable for IP66/67, NEMA 4/4X/6. Figure 2-16: IP68 versions 1 Submersible 2 Buried 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
• Step 1: Apply approx. 50% of max. torque given in table. • Step 2: Apply approx. 80% of max. torque given in table. • Step 3: Apply 100% of max. torque given in table. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
Page 17
12 x 7/8" 1 The torque values also depend on variables (temperature, bolt material, gasket material, lubricants, etc.) outside the control of themanufacturer. Therefore these values should be regarded as indicative only. 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
• Before closing the case of the converter, ensure that all surfaces in contact with the seals are clean. • Position the upper part of the case and tighten the lock ring. • Use the wrench to tighten the ring as shown. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
The device must be grounded in accordance with regulations in order to protect personnel against electric shocks. Figure 3-1: Grounding INFORMATION! Grounding without grounding rings. The flow sensor is equipped with a reference electrode. 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
3 Brown + white + violet cable, used for electrode signals (terminals 1, 2, 3) 4 Drain wires Dimensions of cable inch Figure 3-3: Cable connection at sensor side, standard cable 1 Connect drain wires under screw 2 Connect shielding under clamp www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
Page 21
• Connect the wires as shown in the following table. Wire color Terminal Function Brown Reference electrode White Standard electrode signal Violet Standard electrode signal Blue Field current Green Field current Yellow No function Drain wires Screws Shielding 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en www.krohne.com...
Mounting of the IP68 • For closing the case of the converter in the IP68 housing, please refer to version on page 18. INFORMATION! The instrument now operates with default settings. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
INFORMATION! The instrument now operates with default settings. For configuration of these settings, please refer to the handbook of the WATERFLUX 3070. 4.2.2 IP68 housing (compact version) The output cable has two color coded leads for connecting the external battery. For detailed...
Especially for smaller nominal sizes of the sensor, the converter can be bigger than the • sensor. Note that for other pressure ratings than mentioned, the dimensions may be different. • • For full information on converter dimensions see relevant documentation. www.krohne.com 03/2012 - 7312342300 - QS WATERFLUX 3070 R03 en...
Page 31