8.2
AUTO CALIBRATION FOR ALL OTHER MODELS
• Before entering calibration mode, press [On/Off] to turn the power
on for 1 minute.
Enter calibration mode
• Press [On/Off] to turn the power off.
• Press [Unit] first and then press [On/Off], while keeping the [Unit]
key pressed. The display will show the last calibration value used
(For example, " L – 1000"), press [Tare] to choose between the two
calibration weight values. When you have made your selection press
[Unit] and the display will show the A/D value.
Calibration
• Wait for the stable indicator to be displayed,
• Press [Unit] -the display will show "00SAVE". After 2 seconds, the
display will return to A/D value.
• Place the required weight on the platform. This is the weight that you
selected as your calibration weight value earlier.
• After the stable indicator is displayed press [Unit]
• The display will show "00SAVE". The calibration in complete now.
• Press [On/Off] to turn the power off. Press [On/Off] to turn on the
power again,
• Place full weight (eg.2000g for AQT2000) on the platform to verify
weighing is correct. If not, repeat the calibration procedure

9.0 ERROR MESSAGES

EEEEHI
Overload
EEEELO
A/D value too low
Low battery