Service Manual
Software rev.: 1.71 & above
ADAM EQUIPMENT CO. LTD.
P.N. 4314, Rev. D, December 2005
© Adam Equipment Company 2005

Advertisement

Table of Contents
loading

Summary of Contents for Adam Equipment AZplus Series

  • Page 1 Adam Equipment AZplus SERIES Service Manual Software rev.: 1.71 & above ADAM EQUIPMENT CO. LTD. P.N. 4314, Rev. D, December 2005 © Adam Equipment Company 2005...
  • Page 2: Table Of Contents

    USING “0000” TO ENTER THE CALIBRATION PARAMETER......16 USING “2003” TO ENTER THE SERVICE PARAMETERS........17 F1 -CALIBRATION ....................18 6.2.1 F2 -RESOLUTION ....................18 6.2.2 F3 -CAPACITY....................18 6.2.3 F4 WEIGHT DECIMAL POINT POSITION ............19 6.2.4 F5 -UNIT PRICE/TOTAL PRICE DECIMAL POINT LOCATION......19 6.2.5 © Adam Equipment Company 2005...
  • Page 3 F9 -A/D COUNT ....................21 6.2.9 F10 –ROUNDING OFF THE TOTAL PRICE ............21 6.2.10 WIRING DIAGRAM ....................22 MECHANICAL ASSEMBLY ..................23 10.0 FRONT DISPLAY SCHEMATIC................25 12.0 MAIN SCHEMATIC LAYOUT..................27 13.0 PILLAR DISPLAY SCHEMATIC ................28 14.0 REAR DISPLAY SCHEMATIC ................29 ***** © Adam Equipment Company 2005...
  • Page 4: Introduction

    1.0 INTRODUCTION The AZplus series of digital price-computing scales provide an accurate, fast and versatile method of computing price by weight. These series of scales are EC approved for the use in UK and Europe. For all revisions from 1.27 and above, the Price per kilogram and per 100 gram [Kg 100g] key can be disabled at the factory for the scales to be used in South Africa and Far East.
  • Page 5: Troubleshooting

    Incorrect adjustment of the mechanical stops. Factory calibration has been tampered with- check the ADC counts and re-calibrate. An error “E9” The scale is not stable to set the initial zero. displays on the screen © Adam Equipment Company 2005...
  • Page 6: Installation

    Secure the pillar rod to the socket by using two M4 screws. A black plastic cap is provided to seal © Adam Equipment Company 2005...
  • Page 7: Display Windows

    The window for the Weight has arrows to indicate when the scale is at zero or if the scale is displaying a net weight (if a tare value has been entered). It also indicates when the scale is stable. © Adam Equipment Company 2005...
  • Page 8: Battery Charging

    When the scale is plugged into the main power, the internal battery will be charged. If the LED is green the battery is fully charged. If it is red the battery is nearly discharged and yellow indicates the battery should be charged longer, preferably overnight. © Adam Equipment Company 2005...
  • Page 9: Key Descriptions

    [Kg 100g] It selects the unit price per 100g or per kilogram. NOTE: For AZplus for South Africa and Far East, this selection of unit price per 100g or per kilogram is not applicable. © Adam Equipment Company 2005...
  • Page 10: Operation

    As product is added only the weight of the product will be shown. When the product and container are removed the weight display will show a negative value. If a different container is to be used, remove this negative tare value by pressing the [Tare] once more. © Adam Equipment Company 2005...
  • Page 11: Determining Value Of Product By Weight

    [Kg 100g] key can be disabled at the factory for the scales to be used in South Africa and Far East. This is done by linking the jumper at K5 position on the PCB. © Adam Equipment Company 2005...
  • Page 12: Using A Stored Plu Value For Unit Price

    Display will show “LOAd” “POS 14” ” ” and the Unit Weight value for PLU 14 will be displayed. Press [PLU4]. Wait for few seconds. Display will show “LOAd” “POS 00” ” ” and the Unit Weight value for PLU 4 will be displayed. © Adam Equipment Company 2005...
  • Page 13: Entering Values For The Plu S

    When the Auto Clear is active the Auto Clear indicator (arrow) on the LCD will be ON. Pressing the [Auto Clear] key will turn off the function and the indicator will disappear. Pressing it will turn the function on again. © Adam Equipment Company 2005...
  • Page 14: Kg/100G Key, Price Per Kilogram, Price Per 100G

    Auto sleep after 30 minutes • Press [Tare] to set the value. The scale returns to zero. NOTE: For AZplus for South Africa and Far East, press the [Zero] key to enter Auto Sleep function. © Adam Equipment Company 2005...
  • Page 15: Backlight Operation

    It will be on when a weight is on the platform or when a key is pressed. It will turn off automatically after a few seconds when the weight is zero or after a while the last key was pressed. © Adam Equipment Company 2005...
  • Page 16: Calibration

    5.0 CALIBRATION CALIBRATION PROCEDURE The scales are sealed to prevent unauthorised calibration. Contact Adam Equipment or your supplier for more details. The calibration of the AZplus scale is accomplished by removing the label at the base of the scale which covers a hole through which the PCB can be accessed.
  • Page 17: Calibration Procedure

    The display will show “ ” “ ” if the calibration is OK. Or it will show “ ” “ ” if calibration could not be completed correctly. The scale will return to the parameter menu. © Adam Equipment Company 2005...
  • Page 18: Using "2003" To Enter The Service Parameters

    Weight decimal point location Point 0.00 0.00 Unit Price/Price decimal point Init ZEro Initial Zero Range rE ZEro Re-Zero range SCSivE tArE Successive Tare Enable A/D counts Display the A/D counts roUnd Rounding off the Total Price © Adam Equipment Company 2005...
  • Page 19: F1 -Calibration

    To enter this parameter, press the [Tare] key when “ ” is shown. The display will show the current capacity in 6, 15 or 30 kilograms. Press the [PLU4] key to change the value. Press [Tare] to accept the displayed value. © Adam Equipment Company 2005...
  • Page 20: F4 Weight Decimal Point Position

    The display will show the current decimal point location in the Unit Price and the Total Price windows. Press the [PLU4] key to change the value. Press [Tare] to accept the displayed value. NOTE: The units must match the currency used in the country. Normally 0.00 is shown. © Adam Equipment Company 2005...
  • Page 21: F6 -Initial Zero Range

    The display will show if the successive tare is on or off. Press the [PLU4] key to change the value. Press [Tare] to accept the displayed value. NOTE: Either option is allowed for the AZplus scales. © Adam Equipment Company 2005...
  • Page 22: F9 -A/D Count

    The display will show the current settings. Press the [PLU4] key to change the value and press [Tare] to accept the value. Press [Tare] to return to the PARAMETER menu. Press [Zero] to return to weighing. © Adam Equipment Company 2005...
  • Page 23: Wiring Diagram

    7.0 WIRING DIAGRAM © Adam Equipment Company 2005...
  • Page 24: Mechanical Assembly

    8.0 MECHANICAL ASSEMBLY © Adam Equipment Company 2005...
  • Page 25 PCBA the load cell lower supporter the load cell the load cell upper supporter washer screw front display PCBA upper cover rear display overlay ABS pan SST pan © Adam Equipment Company 2005...
  • Page 26: Front Display Schematic

    10.0 FRONT DISPLAY SCHEMATIC © Adam Equipment Company 2005...
  • Page 27 11.0 MAIN PCB ASSEMBLY © Adam Equipment Company 2005...
  • Page 28: Main Schematic Layout

    12.0 MAIN SCHEMATIC LAYOUT © Adam Equipment Company 2005...
  • Page 29: Pillar Display Schematic

    13.0 PILLAR DISPLAY SCHEMATIC © Adam Equipment Company 2005...
  • Page 30: Rear Display Schematic

    14.0 REAR DISPLAY SCHEMATIC © Adam Equipment Company 2005...
  • Page 31 © Adam Equipment Company 2005...
  • Page 32 © Copyright by Adam Equipment Co. Ltd. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment. Adam Equipment reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.