3.3 Reviewing the Calibration

10. To save the calibration and return to the Reading mode, press
. To exit the calibration without saving it and return to
ENTER
the reading mode, press
1. From the pH Reading mode, press the
2. The display will show the standard number, standard pH, and
temperature (view mV value by pressing
arrow once.
3. The meter will continue to scroll through the standard
information with each press of the up arrow key. When all the
standards have been displayed, press the up arrow key again.
4. The meter will display the slope and offset of the
calibration curve.
5. To exit Cal Review mode, press
or slope information again, press the down arrow.
29
.
EXIT
REVIEW
pH/mV
. To review any standard
EXIT
key.